DentiMax Dream IOX TWAIN - AutoAdvance
- DentiMax TWAIN Driver UI has a tray icon where it can be configured for single shot or multi-shot.
DentiMax Sensor Image Quality Configurations
- DentiMax has a config.ini file with image preprocessing values that can be configured located in : C:\ProgramData\DentiMax\
- Customer should call DentiMax with a clinical provider who makes judgements about IQ.
- Have DentiMax support assist clinical provider to dial in the image quality to to their standards.
- Then, they should replace that config.ini file in every room to make those filters consistent across the practice to replicate their image quality standards to each room.
How to solve miorroring
- C:\ProgramData\DentiMax\
- Open the DentiMaxDS.ini in Notepad
- Change the "FlipY=0" to "1"
- Save and Exit.
How to solve DentiMax TWAIN exit in EzDent-i
Solve an issue in EzDent-i where a user can not interrupt acquisition of the DentiMax TWAIN while having mounting squares were left blank in the sequence or not aquired.
- C:\Program Files (x86)\VATECH\EzDent-i\Setting\

- Open the TwainConfig.ini in Notepad and change "CLOSE_NO_PENDING_IMAGE = ON"