Please use the following search box to quickly locate valuable FAQ support guide.
You can use key words or full statement to search by. Alternatively you can simply browse by Category belowEzRay Air Portable Alarm Code Troubleshooting Guide
EzRay Air Portable has self-diagnostic technology and perform self-diagnosis by monitoring internal signals for evidence of faults. Please follow instructions below.
Alarm Code |
Description |
Troubleshooting Guide |
A06 |
Tube temperature is higher than 149 °F (65 °C) |
Wait 20 ~ 30 min then try again |
A07 |
Exposure switch was pressed while the system power is in the process of turning on |
Release exposure switch and reboot the EzRay Air P |
A08 |
Exposure switch was released too earlier than the exposure time |
Press the exposure switch enough until you hear a beep sound, or the indicator light goes out |
A10 |
Battery Current is lower than 21V |
1. Charge the Battery and test again 2. If this doesn’t work, replace the battery |
A11 |
Exposure switch was pressed while charging the battery |
Unplug battery charger from EzRay Air P and use it |