EzRay Air Portable Alarm Code

Written on Jun 15, 2020

EzRay Air Portable Alarm Code Troubleshooting Guide

 

EzRay Air Portable has self-diagnostic technology and perform self-diagnosis by monitoring internal signals for evidence of faults. Please follow instructions below.

Alarm Code

Description

Troubleshooting Guide

A06

Tube temperature is higher than 149 °F (65 °C)

Wait 20 ~ 30 min then try again

A07

Exposure switch was pressed while the system power is in the process of turning on

Release exposure switch and reboot the EzRay Air P

A08

Exposure switch was released too earlier than the exposure time

Press the exposure switch enough until you hear a beep sound, or the indicator light goes out

A10

Battery Current is lower than 21V

1. Charge the Battery            and test again

2. If this doesn’t work, replace the battery

A11

Exposure switch was pressed while charging the battery

Unplug battery charger from EzRay Air P and use it

 

 

 

 